Super Systems MGA6000 User Manual

Page 29

Advertising
background image

SSi Manual #4563 Rev. D Page 29 Multi-Gas Model 6000

65 flow

span

value

* Register 66 not used

67 Dualpro

process

factor

register

68

LED D2 ON time in millisecs

69

LED D2 OFF time in millisecs

70 Display

polarity

71

Display contrast

72

Backlight brightness

73

Backlight on time; 0 = always ON

74

Web change enable; 0 = disable, 1 = enable

75

Calibration

stage

76

Pump ON delay

77 Sample

OFF

delay

78

Adjust/control update interval

79

sample ON delay time

80

H2 zero gas % times 100 (xx.xx)

81

H2 span gas % times 100 (xx.xx)

82

hydrogen cell calibration request; 1 = zero, 2 = span

83

H2 cell in % times 100 (xx.xx)

84

Help

pointer

85

Sum of Active Cell start up timers

86

Backlight

brightness

output

87 to 95

slave communications status

96

PIC processor communications status (96 - 99)

100

local cooler ON set point

101

local cooler OFF set point

102

auto calibration and/or sequencing

bit 0 = auto cal; 0 = OFF, 1 = ON

bit 1 = sequencing; 0 = OFF, 1 = ON

bit 2 = seq mode; 0 = normal; 1 = specific

bits 4 - 7 Not Used

bits 8 - 13 CH active bit map

bits 14 - 15 Not Used

103

Auto cal interval in minutes 0 = OFF

104

Auto Zero interval in minutes 0 = OFF

105

Auto data display time at end of seq

106

purge time before cal or zero

* Registers 107 to 109 not used

* Selected gas types:

0 = none

1 to 4 = PIC 0 to 3 SPI IR cells respectively

5 = H2 cell via RS232

6 = Lambda probe via RS-485

7 = PIC 2 A/D1 Oxygen cell

8 = PIC 2 A/D3 Ext thermister

9 = PIC 1 A/D3 Ext 0 to 5 volt

10 = PIC 3 A/D0 Ext 0 to 1 volt

11 = PIC 3 A/D1 Ext 0 to 1 volt

12 = PIC 3 A/D3 Ext 0 to 5 volt


Selected Calculation types:

0 = none

1 = IR % carbon (needs CO and CO2 and optionally CH4)

2 = Gas ratio (Gas_1_val/Gas_2_val)