Super Systems MGA6000 User Manual

Page 30

Advertising
background image

SSi Manual #4563 Rev. D Page 30 Multi-Gas Model 6000

3 = Dew point (need to define how calculated)

Message selections:

0 = use name (Gas_x_name),

1 = CO,

2 = CO2,

3 = CH4,

4 = H2,

5 = O2,

6 = CO/CO2 Ratio,

7 = Probe MV,

8 = Probe TC,

9 = Probe %C,

10 = IR %C,

11

=

NH3,

12 = DP,

>= 20 is blank

* Gas 1

110

Selected gas type; 0 = none

111

Reading for gas 1

112

Full scale range for gas 1

113

Decimal place location; low byte = display, hi = source

114

Span gas value for gas 1

115

Message Selection

116

Calcuation Type; 0 = blank, 1 = %, 2 = deg F, 3 = deg C,

* Registers 117 to 119 not used

* Gas 2
120

Selected gas type; 0 = none

121

Reading for gas 2

122

Full scale range for gas 2

123

Decimal place location; low byte = display, hi = source

124

Span gas value for gas 2

125

Message Selection

126

Calcuation Type; 0 = blank, 1 = %, 2 = deg F, 3 = deg C,

* Registers 127 to 129 not used

* Gas 3

130

Selected gas type; 0 = none

131

Reading for gas 3

132

Full scale range for gas 3

133

Decimal place location; low byte = display, hi = source

134

Span gas value for gas 3

135

Message Selection

136

Calcuation Type; 0 = blank, 1 = %, 2 = deg F, 3 = deg C,

* Registers 137 to 139 not used
* Gas 4

140

Selected gas type; 0 = none

141

Reading for gas 4

142

Full scale range for gas 4

143

Decimal place location; low byte = display, hi = source

144

Span gas value for gas 4

145

Message Selection

146

Calculation Type; 0 = blank, 1 = %, 2 = deg F, 3 = deg C,

* Registers 147 to 149 not used

* Gas 5

Advertising