Introduction, Possible configurations, Specifications – Super Systems MGA6000 User Manual

Page 4

Advertising
background image

SSi Manual #4563 Rev. D Page 4 Multi-Gas Model 6000

MGA6000 Operating Instructions

Introduction

The Model MGA6000 is a continuous Multi-Gas Analyzer using NDIR (Non-Dispersive Infra-Red) technology. It is

capable of measuring between one and four gases, depending on how it is configured at the factory prior to shipment.

The unit is designed and manufactured for the atmosphere heat treating industry.

Possible Configurations

CO range:

0.00 to 30.00 % *

CO

2

range:

0.000 to 2.000 %

CO

2

High range:

0.0 to 20.00 %

CH

4

range:

0.00 to 15.00 % *

O

2

range:

0.1 to 25.0%

Hydrogen:

0.0 to 100.0%

Dew point

0 to 80 F

* Note: The CO and CH4 sensors have been calibrated to be most accurate within the ranges as shown

above, however both sensors are capable of measuring gas concentrations of up to 100.00%.

Specifications


Sampling method:

Extraction by internal pump (when necessary)


Accuracy and repeatability:

± 1% of full scale


Flow Meter:

0 to 2 SCFH, mounted on enclosure front and also displayed on-

screen

Pump Operation:

On/Off/Auto Sample


AC Power:

90 to 230 VAC, 50 to 60 Hz, 60 Watts


Communications:

Ethernet,

RS485


4-20 mA Outputs:

Four (4) user-defined


Digital Inputs:

Two (2) for stopping pump and inhibiting COF/PF adjustment


Relays:

One (1) for alarm or control


Operating Temperature: 32

° to 122° F (0° to 50° C)


Dimensions:

Approx. 16” x 14” x 8”


Weight:

Approx.

20

lbs.