Alarms

O2 Heater Open

Figure 8-7. Fault 5, Open Heater

e. Fault 5, Open Heater

Figure 8-7 shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).

Membrane Keypad. When Fault 5 is

detected, the HEATER LED flashes once,
pauses for three seconds, and repeats.

1. Remove power.

2. Remove the electronic assembly per

paragraph 9-4c, Electronic Assembly
Replacement.

3. Using a multimeter, measure the re-

sistance across the terminals of heater
connector, J8.

4. The measurement should be approxi-

mately 72 ohms. If the heater is open,
see paragraph 9-4g, Heater Strut
Replacement.

LOI. When Fault 5 is detected, the LOI dis-

plays the “O2 Heater Open” message.

1. Remove power. Unscrew and remove

the LOI module from the electronic as-
sembly.

2. Perform the diagnostic steps 2 through

4 shown for the membrane keypad.