Instruction Manual

IB-106-340C Rev. 4.1
July 2004

6-6 Startup and Operation with LOI

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

d. SYSTEM/Status

1. Alarms – Diagnostic alarms. See

Section 8, TROUBLESHOOTING.

2. PID Parameter – Displays the line volt-

age, powering the Oxymitter, and infers
the temperature control algorithm being
used to control heater temperature.

3. Reset Device – Device can be reset

here as opposed to re-powering. Cali-
bration parameters will be lost.

e. SYSTEM/Software

This is data regarding the Oxymitter 4000

software version, and errors that may have
occurred.

f. SENSOR

DATA

Displays information about the O

2

cell and

thermocouple.

1. Temperatures

(a) O2 Temp – Indicates the thermo-

couple temperature at the sensing
cell; this should always be 736°C.

(b) O2 Temp Max. – Maximum tem-

perature the cell has seen. (Some
process temperatures can exceed
the 736°C setpoint temperature,
and this will indicate this condition.)

(c) Board Temp – The temperature

inside the Oxymitter electronics
housing (85°C is the max.).

(d) Board Temp Max. – This is the

maximum temperature that the
electronics has experienced over
time.

6-7

LOI INSTALLATION

The LOI connects to the top of the electronic
assembly in the electronics housing. There are
four matching connectors on the back of the LOI
module, Figure 6-5, that allow the LOI to be ori-
ented as desired by the user.

37260055

CONNECTOR

RECEPTACLES

LOI MODULE

REAR VIEW

Figure 6-5. LOI Module Connectors