July 2004

1-12 Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

1-7 SPECIFICATIONS

O

2

Range:

Standard.............................................

0 to 10% O

2

, 0 to 25% O

2

, 0 to 40% O

2

(via HART)

Accuracy ............................................

±0.75% of reading or 0.05% O

2

, whichever is greater

System Response to Calibration Gas

Initial – less than 3 sec., T90 – less than 8 sec.

Temperature Limits:

Process ..............................................

0° to 704°C (32° to 1300°F) up to 1300°C (2400°F) with
optional accessories

Electronics Housing ...........................

-40° to 70°C (-40° to 158°F), ambient

Electronics Package ..........................

-40° to 85°C (-40° to 185°F) – operating temperature of

electronics inside instrument housing, as measured by

a HART communicator or Rosemount Asset Management
Solutions software

Local Operator Interface ...................

-40° to 70°C (-40° to 158°F), ambient

-40° to 85°C (-40° to 185°F), internal [At temperatures

above 70°C (158°F) inside instrument housing, the infrared

keypad will cease to function, but the Oxymitter 4000 will
continue to operate properly.]

Probe Lengths............................................

18 in. (457 mm), 3 ft (0.91 m), 6 ft (1.83 m)

Mounting and Mounting Position ...............

Vertical or horizontal;

a spool piece (P/N 3D39761G02) is available to offset

transmitter housing from hot ductwork.

Materials:

Probe..................................................

Wetted or welded parts - 316L stainless steel (SS)

Non-wetted parts - 304 SS, low-copper aluminum

Electronics Enclosure ........................

Low-copper aluminum

Calibration..................................................

Manual, semi-automatic, or automatic

Calibration Gas Mixtures Recommended..

0.4% O

2

, Balance N

2

8% O

2

, Balance N

2

Calibration Gas Flow .................................

5 scfh (2.5 l/m)

Reference Air .............................................

0.5 scfh (0.25 l/hr), clean, dry, instrument-quality air

(20.95% O

2

), regulated to 5 psi (34 kPa)

Electronics .................................................

NEMA 4X, IP66 with fitting and pipe on reference exhaust

port to clear dry atmosphere

Line Voltage ...............................................

90-250 VAC, 48/62 Hz; 3/4 in.-14 NPT conduit port.

Pollution Degree ........................................

2

Over Voltage Category ..............................

II

Relative Humidity .......................................

5 to 95% (non-condensing)

Signals:

Analog Output/HART .........................

4-20 mA isolated from power supply, 950 ohms

maximum load

Logic I/O.............................................

Two-terminal logic contact configurable as either an alarm

output or as a bi-directional calibration handshake signal to

IMPS 4000 or SPS 4000.

Self-powered (+5 V), in series with 340 ohms
Conduit port — 3/4 in.-14 NPT (for analog output
and logic I/O signal lines)