Figure 8-11. fault 9, high heater temp -13, Hazardous area oxymitter 4000, Alarms o2 temp hi – Emerson Process Management OXYMITTER 4000 User Manual

Page 109: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-13

Hazardous Area Oxymitter 4000

Alarms

O2 Temp Hi

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260027

Figure 8-11. Fault 9, High Heater Temp

i. Fault 9, High Heater Temp

Figure 8-11 shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).

Membrane Keypad. When Fault 9 is

detected, the HEATER LED flashes five
times, pauses for three seconds, and
repeats.

1. If the thermocouple produces a voltage

in excess of approximately 30.7 mV,
the high heater temp alarm activates.

2. The 4-20 mA signal returns to the

default value (4 or 20 mA).

3. This alarm is self-clearing. When tem-

perature control is restored and the
thermocouple voltage returns to the
normal range, the alarm clears.

4. If the temperature continues to rise, the

next alarm will be the high high heater
temp alarm.

LOI. When Fault 9 is detected, the LOI

displays the “O2 Temp Hi” message. Refer
to the comments and procedures in para-
graphs 1 through 4 above.