Hazardous area oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 User Manual

Page 65

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

Configuration with Keypad 3-3

Hazardous Area Oxymitter 4000

1. Access TP5 and TP6 next to the mem-

brane keypad. Attach a multimeter
across TP5 and TP6. The calibration
and process gases can now be moni-
tored. Pressing the INC or DEC once
will cause the output to switch from the
process gas to the calibration gas.
Pressing INC or DEC a second time
will increase or decrease the calibra-
tion gas parameter. If the keys have
been inactive for one minute, the

output reverts to the process gas.
When a calibration has been initiated,
the value at TP5 and TP6 is the %O

2

seen by the cell. Oxygen levels, as
seen on the multimeter, are:

8.0% O

2

= 8.0 VDC

0.4% O

2

= 0.4 VDC

2. HART/AMS.

3. Model 751. The loop-driven LCD

display.

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mV -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

SW1

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

4-20 mA IS

INTERNALLY

POWERED (DEFAULT)

4-20 mA REQUIRES
EXTERNAL POWER

HART

0 TO 10% O

2

3.5 mA

220 V

115 V

0 TO 25% O

2

LOCAL

21.6 mA

DEFAULT

POSITION

(EX-FACTORY)

3.5 mA/21.6 mA:

0 TO 25% O :

2

0 TO 10% O /

2

LOCAL:

HART: O RANGE SET BY HART/AMS

(FROM 0 TO 40% O )
O RANGE SET BY POS 2

O RANGE
WHEN ALARM EXISTS, OR ON
POWER-UP, OUTPUT CURRENT
GOES TO THIS VALUE

2

2

2

2

1

2
3
4

37270001

ON

OFF

NOTE:

THE 115 V OPTION AT
SWITCH SW2 POSITION 4
IS ACTIVE ONLY WHEN
THE HEATER VOLTAGE
OPTION IS SET TO MANUAL
IN THE SOFTWARE
(AUTO TUNE = NO).

Figure 3-2. Defaults – Hazardous Area Oxymitter 4000 with Membrane Keypad