Defining a timed calibration via hart -9, Hazardous area oxymitter 4000 – Emerson Process Management OXYMITTER 4000 User Manual

Page 95

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

HART/AMS 7-9

Hazardous Area Oxymitter 4000

7-8

DEFINING A TIMED CALIBRATION VIA
HART

Use the following procedure to specify a time
interval (in hours) at which the Hazardous Area
Oxymitter 4000 will be automatically calibrated.
If necessary, use the menu tree in Figure 7-3
(sheet 2 of 3) for reference.

NOTE

To select a menu item, either use the
up and down arrow keys to scroll to
the menu item and press the right
arrow key or use the number keypad
to select the menu item number. To
return to a preceding menu, press the
left arrow key.

a. From the DEVICE SETUP screen, select

DETAILED SETUP.

b. From the DETAILED SETUP screen, select

O

2

CALIBRATION.

c. From the O

2

CALIBRATION screen, select

menu item 6, CAL MODE. Set the CAL
MODE to AUTO.

d. Return to the O

2

CALIBRATION screen and

select menu item 7, CAL INTRVL.

e. At the prompt, input a time interval

(in hours) at which an automatic calibration
will occur; then press ENTER.

7