Figure 8-10. fault 8, low heater temp -12, Hazardous area oxymitter 4000, Alarms o2 temp low – Emerson Process Management OXYMITTER 4000 User Manual

Page 108: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1
July 2004

8-12 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260026

Alarms

O2 Temp Low

Figure 8-10. Fault 8, Low Heater Temp

h. Fault 8, Low Heater Temp

Figure 8-10 shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).

Membrane Keypad. When Fault 8 is

detected, the HEATER LED flashes four
times, pauses for three seconds, and
repeats.

1. The low heater temperature alarm is

active when the thermocouple reading
has dropped below 28.6 mV.

2. If the thermocouple reading continues

to ramp downward for one minute and
does not return to the temperature set
point of approximately 29.3 mV, then
an Open Heater fault will be displayed.

3. Power down the electronics. Remove

the electronic assembly per paragraph
9-4c, Electronic Assembly Replace-
ment. Using a multimeter, measure the
resistance across the terminals of
heater connector, J8.

4. If the heater is good, the reading will be

approximately 70 ohms. If the heater is
open, see paragraph 9-4g, Heater Strut
Replacement.

LOI. When Fault 8 is detected, the LOI

displays the “O2 Temp Low” message.
Refer to the comments and procedures in
paragraphs 1 through 4 above.