Figure 8-12. fault 10, high cell mv -14, Hazardous area oxymitter 4000, Alarms o2 cell open – Emerson Process Management OXYMITTER 4000 User Manual

Page 110: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1
July 2004

8-14 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

Alarms

O2 Cell Open

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260028

Figure 8-12. Fault 10, High Cell mV

j. Fault 10, High Cell mV

Figure 8-12 shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view). The upper view also
shows J1 and test points TP1 through TP6,
located on the microprocessor board, below
the membrane keypad or the LOI module.

Membrane Keypad. When Fault 10 is

detected, the O2 CELL flashes once,
pauses for three seconds, and repeats.

1. Using a multimeter, measure across

TP1+ to TP2-. If you measure 204 mV
to 1 volt DC, the cell reading is due to
high combustibles. This is a self-
clearing alarm, once the combustible
conditions go away. If you measure
1.2 VDC, the cell wires, either orange
or green, have become detached from
the input.

2. One possible cause is connector J1.

The orange or green wire has come
loose from the crimped connection.

3. The platinum pad could also be at fault.

The pad could have broken free from
the back of the cell.

4. Replace heater strut per paragraph

9-4g, Heater Strut Replacement. If
necessary, replace the cell and flange
assembly per paragraph 9-4h, Cell
Replacement.

LOI. When Fault 10 is detected, the LOI

displays the “O2 Cell Open” message.

1. Remove power. Unscrew and remove

the LOI module from the electronic
assembly.

2. Reconnect power to the Oxymitter

4000.

3. Perform the diagnostic steps 1 through

4 shown for the membrane keypad.

Advertising