Figure 8-17. fault 15, last calibration failed -19, Hazardous area oxymitter 4000, Alarms calib failed – Emerson Process Management OXYMITTER 4000 User Manual

Page 115: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

Troubleshooting 8-19

Hazardous Area Oxymitter 4000

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

37260033

Alarms

Calib Failed

LOI

KEYPAD

Figure 8-17. Fault 15, Last Calibration Failed

o. Fault 15, Last Calibration Failed

Figure 8-17 shows the electronic assembly

for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).

Membrane Keypad. When Fault 15 is

detected, the CALIBRATION LED flashes
three times, pauses for three seconds, and
repeats.

1. The last calibration failed alarm acti-

vates when the slope and constant
values calculated are out of range and
the unit reverts to using the previous
calibration values.

2. The cell should be replaced. See para-

graph 9-4h, Cell Replacement, for cell
replacement instructions.

LOI. When Fault 15 is detected, the LOI

displays the “Calib Failed” message. Refer
to the comments in paragraphs 1 and 2
above.

8