Imps 4000 connections -15, Sps 4000 connections -15, To figure 2-11 – Emerson Process Management OXYMITTER 4000 User Manual

Page 61: Hazardous area oxymitter 4000, Instruction manual, Figure 2-11. air set, plant air connection, And high gas - 8% o

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-15

Hazardous Area Oxymitter 4000

37270008

REF AIR SET

263C152G01

1

FLOWMETER

0.006-0.057 SCMH (0.2-2.0 SCFH) 771B635H02

2

PRESSURE GAGE

0-103 KPA GAGE (0-15 PSIG)

275431-006

3

COMBINATION FILTER-REG. 0-207 KPA GAGE (0-30 PSIG)

4505C21G01

NOTE: DIMENSIONS ARE IN MILLIMETERS

WITH INCHES IN PARENTHESES. ALL

PIPING SPECIFIED IN U.S. STANDARDS.

1

2

3

122.17 (4.81)

FLOW SET

POINT KNOB

0.125-27 NPT

OUTLET

30.22

(1.19)

254

(10.0)

DRAIN

VALVE

79.25 (

MAX

3.12)

215.90

(8.50)

MAX

50.80

(2.0)

MOUNTING

HOLES

81.03

(

) LG

FOR

7.92 (

)

DIA.

THROUGH

BOLTS

3.19

0.312

38.10

(1.50)

57.15 (2.250)

SCHEMATIC HOOKUP FOR REFERENCE

AIR SUPPLY TO HAZARDOUS AREA

OXYMITTER 4000 PROBE HEAD.

OUTLET

0.25-18

NPT

INLET

INSTRUMENT AIR

69-1551 KPA GAGE

(10-225 PSIG) MAX.

0.250 IN. OR 6 MM O.D. TUBING

(SUPPLIED BY CUSTOMER)

500 VA

SERIAL NO.
TAG NO.

OXYMITTER 4000

WATTS:

VOLTS:

FUSE:

LINE

OUTPUT:

Rosemount Analytical Inc.
Orrville, OH 44667-0901

85-264 VAC 48-62 Hz

TM

800-433-6076

4-20 mA

R

5 Amps

TM

HART

SMART FAMILY

REF

2

3

1

Figure 2-11. Air Set, Plant Air Connection

500 VA

R

SERIAL NO.
TAG NO.

OXYMITTER 4000

VOLTS:

WATTS:

OUTPUT:

LINE FUSE:

Rosemount Analytical Inc.
Orrville, OH 44667-0901

85-264 VAC 48-62 Hz

TM

800-433-6076

4-20 mA

R

5 Amps

TM

HART

SMART FAMILY

CALIBRATION GAS

REFERENCE AIR

26170025

Figure 2-12. Hazardous Area Oxymitter 4000

Gas Connections

Do not use 100% nitrogen as a low gas
(zero gas). It is suggested that gas for
the low (zero) be between 0.4% and
2.0% O

2

. Do not use gases with hydro-

carbon concentrations of more than 40
parts per million. Failure to use proper
gases will result in erroneous
readings.

b. Calibration

Gas

Two calibration gas concentrations are used

with the Hazardous Area Oxymitter 4000,

Low Gas - 0.4% O

2

and High Gas - 8% O

2

.

See Figure 2-12 for the Hazardous Area
Oxymitter 4000 connections.

If using an IMPS 4000 or an SPS 4000,
install it in a non-hazardous, explosive-
free environment.

2-5

IMPS 4000 CONNECTIONS

Ensure the IMPS 4000 is installed in a safe
(non-hazardous, explosive-free) area and verify
the wiring and pneumatic connections per the
IMPS 4000 Intelligent Multi-probe Test Gas
Sequencer Instruction Bulletin.

2-6

SPS 4000 CONNECTIONS

Ensure the SPS 4000 is installed in a safe (non-
hazardous, explosive-free) area and verify the
wiring and pneumatic connections per the SPS
4000 Single Probe Autocalibration Sequencer
Instruction Bulletin.

2