Hart communicator pc connections -3, Off-line and on-line operations -3, Hazardous area oxymitter 4000 – Emerson Process Management OXYMITTER 4000 User Manual

Page 89: Instruction manual

Advertising
background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

HART/AMS 7-3

Hazardous Area Oxymitter 4000

LOOP CONNECTORS

USE INTERFACE
00275 0013 ONLY

SERIAL PORT & BATTERY
CHARGER MUST
NOT BE USED IN
HAZARDOUS AREAS

SERIAL PORT

AC L1

AC N

+

+

-

-

4-20

TERMINAL BLOCK

LOOP CONNECTORS

4-20 mA SIGNAL LINE

HART

COMMUNICATOR

RL < 250

ANALOG OUTPUT DEVICE

250 OHM

LOAD

RESISTOR

(NOTE)

HART COMMUNICATOR

REAR PANEL

NOTE: THE SIGNAL LOOP MUST BE BROKEN

TO INSERT THE OPTIONAL 250 OHM
LOAD RESISTOR.

23230002

Figure 7-2. Signal Line Connections, < 250 Ohms Load Resistance

7-3

HART COMMUNICATOR PC
CONNECTIONS

There is an option to interface the HART Com-
municator with a personal computer. Load the
designated AMS software into the PC. Then link
the HART Communicator to the PC using the
interface PC adapter that connects to the serial
port (on the communicator rear panel).

Refer to the proper HART Communicator docu-
mentation in regard to the PC interface option.

7-4

OFF-LINE AND ON-LINE OPERATIONS

The HART Communicator can be operated both
off-line and on-line.

a. Off-line operations are those in which the

communicator is not connected to the Haz-
ardous Area Oxymitter 4000. Off-line op-
erations can include interfacing the HART
Communicator with a PC (refer to applica-
ble HART documentation regarding
HART/PC applications.

b. In the on-line mode, the communicator is

connected to the 4-20 mA analog output
signal line. The communicator is connected
in parallel to the Hazardous Area Oxymitter
4000 or in parallel to the 250 ohm load
resistor.

NOTE

If the HART Communicator is turned
on while connected to the 4-20 mA
analog output signal line, an undefined
status indication appears while the
communicator warms up. Wait until
the warmup period ends to continue.

c. The opening menu displayed on the LCD is

different for on-line and off-line operations.
When powering up a disconnected (off-line)
communicator, the LCD will display the
Main Menu. When powering up a connected
(on-line) communicator, the LCD will display
the On-line Menu. Refer to the HART
Communicator manual for detailed menu
information.