LogTag Recorders LogTag Analyzer Temperature Recorder User Manual

Page 31

Advertising
background image

Chapter 4 Preparing LogTag(s) for use

31

Start method

This is used to determine how/when the LogTag will start recording

its next set of readings.
 Push button start

When the 'Push button start' option is selected, the LogTag will

start recording when the START MARK button has been pressed

and held until both the OK and Alert lights alternately flash on
and off. If pre-start recording is enabled, the LogTag will start

logging pre-start readings immediately (see section "

Pre start

data collection

on page 42").

 Date/Time start

When the 'Date/Time start' option is selected, the LogTag will
automatically start recording at the date/time defined in the

subsequent field. Pressing the START MARK button prior to the

start date/time will have no effect. There will be no pre-start

readings recorded prior to the start date/time.
Note: When using the Date/Time start option, the earliest start

time that can be selected is two minutes after the current time.

This delay allows you to complete the remaining configuration
options and gives the software enough time to configure the

LogTag before it starts.

 Push button start from hibernation

When the 'Push button start from hibernation' option is
selected, the LogTag will start as if programmed for 'Push

button start'. The difference is that once the configuration is

successfully downloaded to the LogTag it will be placed into

hibernation. When the START MARK button has been pressed

and held, the LogTag will wake from hibernation and start

recording with the parameters as configured.

 Please note in this mode the real time clock inside the logger is

not active. It is therefore vital that the computer on which the

unit is downloaded is set to the correct time.
Please see

Prolonging battery life

(on page 45) for more

information on Hibernation.

Not every LogTag model can support every start mechanism. As a

result, specific options may not be be listed when configuring a
LogTag.

Enable pre-start logging Pre-start logging is a feature that protects the recording in the

event of a failure to actually press the START MARK button to
initiate the recording. It is an insurance policy against the loss of

critical data. This is almost always checked for transit monitoring

applications. This option is further explained in the section "

Pre

start data collection

on page 42".

Record readings

continuously, overwrite

oldest when memory

full

This is what is called a "data wrap" in which the recorder just keeps

writing data to memory. It is a "moving window" of data. For

example, this is useful for fixed base MKT monitoring with the

interval set to one year, so that the current "yearly MKT" can

always be in the memory of the LogTag. If this option is selected,
the user will be "opted out" of the next two options. This option is

further explained in the section "

Continuous operation

(on page

43)".