LogTag Recorders LogTag Analyzer Temperature Recorder User Manual

Page 38

Advertising
background image

38

LogTag Analyzer User Guide (2.0)

Function

Purpose

Readings above X

Readings below Y
(inclusive)

In addition to entering the threshold levels, tick the boxes next

to "Readings above" or "Readings below" to enable the alert
conditions for only the upper or only the lower threshold

settings, or for both. If both boxes are ticked, the behaviour is

the same as for the standard alert settings. The charts will

display alert readings according to the settings made in the

Charts

(on page 90) section of the 'Options' Screen.

After X consecutive alert

readings (W minutes)

Readings above

Readings below

Tick the boxes next to "Readings above" or "Readings below" to

establish whether only readings above and including the upper

threshold or only readings below and including the lower

threshold should count towards the consecutive alert readings. If
both boxes are ticked, the behaviour is the same as for the

standard alert settings. If the Alert Indicator is disabled for either

the upper or the lower alert, the respective box for the

consecutive alert is grayed out and cannot be ticked.

After X alert readings have

occurred (W minutes)

Readings above

Readings below

Tick the boxes next to "Readings above" or "Readings below" to

establish whether only readings above and including the upper

threshold or only readings below and including the lower

threshold should count towards the total number of alert

readings. If both boxes are ticked, the behaviour is the same as
for the standard alert settings. If the Alert Indicator is disabled

for either the upper or the lower alert, the respective box for the

alert total is grayed out and cannot be ticked.

The advanced settings are displayed automatically without the need to press

if an advanced configuration is already found in the logger to be configured. If you do not
wish to use the advanced settings, press

, and the

standard Alert P

rocessing

(on page 34) configuration screen is displayed.

Example:

You wish to configure a LogTag so an alert is displayed if 5 continuous readings are
out-of-range or 7 total readings are out-of-range, but only if the temperature goes to
20ºC and above. You wish to see in LogTag Analyzer if the product has gone to 0ºC and
below, but you do not wish to trigger an alert for this condition. Alert parameters should
be set as follows: