36

LogTag Analyzer User Guide (2.0)

Example:

You wish to configured a LogTag so alert conditions are indicated if 5 continuous readings
are out-of-range or 7 total readings are out-of-range. Alert parameters should be set as
follows:


This table shows how different conditions would result in the Alert Indicator flashing or
not:

What the LogTag recorded

What the LogTag indicated

2 readings out-of-range, return to normal,

2 readings out-of-range, return to normal,

2 readings out-of-range, return to normal

Neither alert condition is met.

The maximum number of consecutive

out-of-spec readings is only 2 (5 required.)
The total out-of-range events equals 2 + 2 + 2

= 6. The alert condition is 7 total, so the alert

does not go off.

No Flashing Red Alert Indicator

4 readings out-of-range, return to normal,

4 readings out-of-range

The first alert condition is not met, as the the

maximum number of consecutive alert condition

readings is only 4 (5 required.)

The second alert condition however IS met
since the total alert condition readings equals 4

+ 4 = 8. The alert condition is 7 total. so the

alert DOES go off.

Flashing Red Alert Indicator

4 readings out-of-range, return to normal,

2 readings out-of-range

Neither alert condition is met.

The maximum number of consecutive alert

condition readings is only 4 (5 required.)

The total number of alert condition readings
equals 4 + 2 = 6. The alert condition is 7 total,

so the alert does not go off.

No Flashing Red Alert Indicator