LogTag Analyzer User Guide (2.0)

n

the number of data points to include in the calculation

t

i

a data point to include in the calculation, in degrees
Celsius

For example, if there were 5 data points to calculate the mean kinetic temperature of and
they were, 12.3, 15.9, 16.2, 14.7 and 14.9 degrees Celsius, then the mean kinetic
temperature would be 14.9 degrees Celsius using the default ΔH.