Diagnostic Mode

5-12

Configuring the Power-up Defaults Menu

CO

2

Test

The CO

2

Test Screen is provided to facilitate troubleshooting problems and performing

verification testing to ensure the measuring results are accurate. See “CO2 Performance” on
page 4-29
for i
nformation on how to perform these tests.

Warning

Warning

A FilterLine, connected to the monitor, should never be applied to a human being while

the monitor is in Diagnostic Mode. Injury could result.

DIAGNOSTIC MENU

CO2 TEST

CO2 Pressure in mmHg (ambient/cell): 745/740
Pump Voltage (Volts): 3.0
CO2 (mmHg): 0.00

Status:

Set Barometric Pressure (mmHg): 745
Flow Rate Check and Calibration:
CO2 Calibration: 07/16/2002 15:13
Pump Op Time: 58

Return