Disassembly Procedures

7-10

Disassembly

Rear Case

Disassembly

You must disassemble the rear case to remove the following modules:

Main PCB

Patient Monitoring I/O

SpO

2

CO

2

NBP

Communications

Power Supply

Caution

Before you remove any module, make sure you are wearing ESD protection and you are

working in a grounded environment.

To disassembly the rear case:

Step

Action

1

Verify that the battery has been removed. If not see, “Removing the Battery on
page 7-4
.

2

Verify that the recorder has been removed. If not see, “Removing the Optional,
External Recorder on page 7-6
. If there is no recorder, unsnap the recorder door
using a flathead screwdriver (see figure below).