Performance Procedures

Testing the Monitor

4-35

Calibration

Verification

Reset Pump

Operating

Time Counters

If the pump in the CO2 module is replaced, the Pump Operating Time counter should be reset
to start counting operating time for the new pump.

Note

When the Pump Op Time has been reset a "CO2 Equipment Malfunction" INOP will be
generated on restarting the monitor. To clear this INOP you must perform a flow check and
store the flow in Diagnostic Mode (select Store Flow).

Step

Action

1

Reopen the 5% gas valve and allow the value to stabilize.

2

Check that the value displayed on the monitor is correct and within the tolerance
(see Step 2 in above section).

3

Disconnect the 5% calibration gas and connect the 10% calibration gas.

4

Open the valve on the flow regulator to allow 10% CO

2

gas to flow into the

monitor. Allow the value to stabilize.

5

Check that the value displayed on the monitor (X7) is correct and within the
tolerance (see Step 6 above). If one or both values are not within tolerances, the
CO

2

module must be exchanged.

Step

Action

1

In the CO2 Test menu, select Pump Op Time.

2

Select Reset to Zero.