Patient Safety Tests

4-42

Testing the Monitor

Enclosure

Leakage

Current

This test is compliance with EN60601-1 (Enclosure Leakage Current). Test at 110% of the
nominal line voltage.

The analyzer leakage current indication must not exceed the values listed below.

Table 4-5. Enclosure Leakage Current

EN60601-1 (1990 + A1, A2, A11, A12, A13) and UL2601 (2

nd

Ed. 1997) US Deviations

What to record on your service record:

S(3): P or S(3): F

Step

Action

1

Connect the AC mains power cord to the analyzer as recommended by the
analyzer operating instructions.

2

Using the appropriate test cable, connect the analyzer to either of the screws on
the back of the monitor, next to the handle.

3

Turn the monitor on.

4

Perform the test as recommended by the analyzer operating instructions.

Test Condition

Polarity

Allowable Leakage Current

Normal

Normal

100 µA

Reversed

100 µA

S.F.C

1

Open

Protective Earth

Normal

300 µA

Reversed

300 µA