Performance Procedures

Testing the Monitor

4-27

SpO

2

Performance

SpO

2

testing includes the following tests:

Dynamic Operating Range

LED Excitation Test

Tools Needed for SpO

2

Tests:

Clinical Dynamics Corp - SmartSat simulator (with Nellcor simulator cable)

SpO

2

adapter cable (M1943A)

SpO

2

reusable sensor, adult finger (M1191A)

Dynamic

Operating

Range Test

The following test sequence verifies proper monitor operation over a range of input signals.

Table 4-1. Settings and Monitor Indications

An * indicates values that produce an alarm. Press the Alarm Silence button to temporarily
silence the audible alarm.

a. For the pulse rate setting of 201 bpm, the pulse rate tolerance of 195 to 207 bpm is greater than ±3 bpm accuracy

specification on the monitor, due to the performance characteristics of the simulator.

Step

Action

1

Connect SpO

2

simulator cable to the SpO

2

adapter cable. Connect the cable to

the SpO

2

patient monitoring input connector.

2

Set the simulator as indicated in the table below. Verify that the monitor readings
are within the indicated tolerances. Allow the monitor several seconds to
stabilize the readings.

3

Turn the monitor Off.

Simulator Settings

Monitor Indications

SpO

2

%

Pulse Rate

Ambient
Light Freq

Ambient
Light Ac
Level

Modulation

SpO

2

Pulse Rate

81

36

120

200

0.50%

79-83*

33-39*

81

112

120

200

5.00%

79-83*

109-115

81

201

a

120

10

0.50%

79-83*

195-207*

81

201

a

120

200

5.00%

79-83*

195-207*