About the RS-232 Interface

10-4

RS-232 Interface

Pin connections for the 9-pin connector are as follows:

Nurse-Call

Pin 9 of the RS-232 serial interface connector provides an Alarm Out signal. Any time there is
any alarm condition active in the monitor, and if the Nurse Call Signal option in the Setup
Menu
is On, pin 9 goes to plus RS-232 level voltage (> +5 VDC). The alarm delay shall be
< 0.5 seconds.

If in the Setup Menu, the Nurse Call Signal option is Off, pin 9 is always at the minus
RS-232 level voltage. In order to make use of the Alarm Out signal, pin 9 should be
connected to a high-impedance circuit (>1000) and protected against transient voltages.

Table 10-1. RS-232 Serial Interface Connections

Pin #

Signal

Direction

1

not used

2

Rx data

input

3

Tx data

output

4

DTR

output

5

Signal Ground

input/output

6

DSR

input

7

RTS

output

8

CTS

input

9

Alarm Out

output