Functionality Assurance Tests

Testing the Monitor

4-13

Step

Action

1

Connect SpO

2

simulator cable to the SpO

2

adapter cable. Connect the cable to

the SpO

2

patient monitoring input connector.

2

Set the simulator as follows:

3

a. Power the monitor On. Verify that the values are within the following

tolerances:

– Oxygen Saturation Range 79% to 83%
– Heart Rate Range 67 to 73 bpm

4

Press the Alarm Silence button to temporarily silence the audible alarm.

5

Verify that the heart rate tone source, found in the Heart Rate Menu, is set to
SpO

2

.

6

Press the Volume button on the monitor’s front panel. Within 3 seconds of
having pressed the button, rotate the navigation wheel clockwise to verify that
the beeping heart rate tone sound level increases.

7

Wait 3 seconds. Press the Volume button and rotate the wheel counter-clockwise
and verify that the beeping heart rate tone decreases until it is no longer audible.
Rotate the wheel clockwise to return the beep volume to a comfortable level.

After 3 seconds with no wheel activity, the volume adjust function terminates.

Item

Setting

Oximeter

Nellcor

SpO

2

%

81

BPM

70

Pulse Mod

5.00%