Site preparation guidelines, Warning, cautions and safety precautions, Patient safety – Philips 862478 User Manual

Page 31: Patient leakage current, Patient safety -3 patient leakage current -3

Advertising
background image

Site Preparation Guidelines

Site Preparations

2-3

Site Preparation Guidelines

Check that the environment is reasonably free from dust, vibration, corrosive or
explosive gases and is within specified temperature and humidity ranges. See
Chapter 12, “Specifications”

Position the monitor so that the display is clearly visible and the controls are easy to
reach.

Warning, Cautions and Safety Precautions

Patient

Safety

To better secure patient safety, become familiar with the details in
Chapter 12, “Specifications.”

Patient

Leakage

Current

The patient leakage current is <= 50µA @ 250 V. The monitor has patient input connectors
(Type CF) that are protected against the effects of defibrillation and electrosurgery.

You will find the following heart symbol on the side of the monitor where the patient
monitoring input connectors are located. The symbol indicates that the monitor is Type CF
and is designed to have special protection against electric shocks (particularly regarding
allowable leakage currents, having a F-Type applied part, according to the standards IEC
60601-1/EN60601-1/CSAC22.2 601.1/UL 2601-1), and is defibrillator proof.