Disassembly Procedures
7-12
Disassembly
Removing the
Main PCB
Module
Step
Action
1
Unscrew the PCB module by loosening the following 3 screws:
2
Gently pull the module towards you.
3
After you have removed the Main PCB module, you should have access to the
CO
2
, SpO
2
, NBP pump, and the Patient Monitoring I/O modules.
Screw #1
Screw #2
Screw #3
SpO
2
Module
Patient
Monitoring
I/O Module
CO
2
Module
NBP Pump