Disassembly Procedures

7-12

Disassembly

Removing the

Main PCB

Module

Step

Action

1

Unscrew the PCB module by loosening the following 3 screws:

2

Gently pull the module towards you.

3

After you have removed the Main PCB module, you should have access to the
CO

2

, SpO

2

, NBP pump, and the Patient Monitoring I/O modules.

Screw #1

Screw #2

Screw #3

SpO

2

Module

Patient
Monitoring
I/O Module

CO

2

Module

NBP Pump