Part 1 Troubleshooting Checklist
Troubleshooting
6-5
The following two steps are to make sure that the monitor’s AC power supply and battery
work correctly.
What To Do If The Monitor Cannot Be Switched On, AC Powered:
What To Do If The Monitor Cannot Be Switched On, Battery Powered:
Step
Action
1
Ensure the battery has had ample time to charge (at least 15 minutes). A dead
battery will prevent the monitor from powering On.
2
Check AC fuses in the AC receptacle.
3
Check keypad connector or replace keypad. See
for instructions.
4
If monitor still cannot be switched On, replace the Main PCB module. See
“Removing the Main PCB Module” on 7-12
for instructions.
Step
Action
1
Plug the monitor into an AC power source.
2
Ensure the battery has had ample time to charge (at least 15 minutes). A dead
battery will prevent the monitor from powering On.
3
Check AC fuses in the AC receptacle.
4
Check keypad connector or replace keypad. See
for instructions.
5
If monitor still cannot be switched On, replace the Main PCB module. See
“Removing the Main PCB Module” on 7-12
for instructions.