Philips 862478 User Manual

Page 115

Advertising
background image

Part 1 Troubleshooting Checklist

Troubleshooting

6-5

The following two steps are to make sure that the monitor’s AC power supply and battery
work correctly.

What To Do If The Monitor Cannot Be Switched On, AC Powered:

What To Do If The Monitor Cannot Be Switched On, Battery Powered:

Step

Action

1

Ensure the battery has had ample time to charge (at least 15 minutes). A dead
battery will prevent the monitor from powering On.

2

Check AC fuses in the AC receptacle.

3

Check keypad connector or replace keypad. See

“Removing the Keypad on

the Front Panel” on 7-6

for instructions.

4

If monitor still cannot be switched On, replace the Main PCB module. See

“Removing the Main PCB Module” on 7-12

for instructions.

Step

Action

1

Plug the monitor into an AC power source.

2

Ensure the battery has had ample time to charge (at least 15 minutes). A dead
battery will prevent the monitor from powering On.

3

Check AC fuses in the AC receptacle.

4

Check keypad connector or replace keypad. See

“Removing the Keypad on

the Front Panel” on 7-6

for instructions.

5

If monitor still cannot be switched On, replace the Main PCB module. See

“Removing the Main PCB Module” on 7-12

for instructions.

Advertising
This manual is related to the following products: