Part 2 Isolating and Solving Monitor Problems

6-10

Troubleshooting

No sound from monitor

Tone is switched off or low.

1. Turn the volume up by pressing the

Volume button.

2. Power-cycle monitor. Tone should

sound when monitor boots up.

3. Check cable connections to the

speaker.

Incorrect Time Stamps

Invalid data stored showing
wrong times for data

The monitor’s real time clock is
defective.

Power the monitor On and then reset the
time and date. See the C3 Instructions for
Use
guide for instructions on how to set
the date and time.

Cannot reprogram the time
and date correctly

Main PCB module malfunction.

Replace the Main PCB module according
to the instructions in

“Removing the

Main PCB Module” on 7-12

.

Fluid dripping from the
monitor

LCD Display damaged or cracked.

Replace LCD Display.

Decrease in light intensity on
display.

Brightness decreases from
top to bottom or bottom to
top.

Backlight tube is worn out.

Replace backlight tube according to
instructions in “Replacing the Backlight
Tube” on page 7-9
.

No tone alarms

Speaker will not work

Audible alarms are not switched
on.

Cable disconnected.

Defective speaker.

Make sure audible alarms are switched on.

Reconnect cable.

Replace speaker.

The wheel is rotated, no
highlight appears on the
display screen, and/or the
monitor does not respond to
wheel presses.

Main PCB module malfunction.

Replace the Main PCB module according
to the instructions in

“Removing the

Main PCB Module” on 7-12

.

Symptom

Cause of Failure

Remedy