Recommendations for maintenance frequency, Recommendations for maintenance frequency -4 – Philips 862478 User Manual

Page 42

Advertising
background image

Recommendations for Maintenance Frequency

3-4

Maintaining the Monitor

Recommendations for Maintenance Frequency

The following table is the recommended maintenance schedule for your C3 monitor. The
maintenance checklist appears in the next section of this chapter. The checklist may be
photocopied and should be completed by the maintainer. It should be filed for future
reference. The owner of the equipment is responsible for the performance of the maintenance
activities with the schedule below.

Maintenance

Frequency

Procedure

Inspect the monitor, cables
and cords

Daily

“Inspecting the Monitor”on page 3-6
and “Inspecting the Cables and
Cords”on page 3-6

Cleaning

As needed

“Cleaning”on page 3-7

Safety checks according to
IEC 60601-1

At least once every 2 years,
after any repairs, if the
monitor has been dropped
or as needed.

“Patient Safety Tests”on page 4-39

Functional testing
procedures

When functional defects in
the measurements are
suspected.

“Functionality Assurance”on page 4-3

Performance Assurance

At least once every year, or
as needed (if you suspect
defects in the
measurements).

“Performance Assurance Test”on page
4-11

Replace backlight

20,000 hours (about 4
years) of continuous use, or
as needed.

“Replacing the Backlight Tube”on
page 7-9

NBP Maintenance

Every year, or as needed, if
the monitor has been
opened, or you suspect
defects in the measurement.

“Preventative Maintenance Tests”on
page 4-14

Battery Maintenance

Every two years

“Battery Maintenance”on page 3-7

Battery Conditioning

Approximately every 200
deep discharges.

“Conditioning a Battery”on page 3-10