Test reporting, Test reporting -5 – Philips 862478 User Manual

Page 55

Advertising
background image

Test Reporting

Testing the Monitor

4-5

Test Reporting

The following table shows what must be recorded on the Service Record after completing the
tests in this chapter.

Where P = Pass, F = Fail and X is the measured value as defined in the tests described in this
chapter.

Test

What to Record

Functionality Assurance Tests

Visual

V:P or V:F

Power On

PO:P or PO:F

Preventive Maintenance Tests

P NBP

PN:P/X1/X2/X3/X4/X5 or
PN:F/X1/X2/X3/X4/X5
See page 4-15 for details

Performance Procedures

CO

2

Performance

PCO

2

: P/X1/X2/X3/X4/X5/X6/X7 or

PCO

2

: F/X1/X2/X3/X4/X5/X6/X7

See page 4-30 for details

Patient Safety Tests

Safety(1) Ground Integrity

S(1):P or S(1):F

Safety(2) Earth Leakage Current

S(2):P or S(2):F

Safety(3) Enclosure Leakage Current

S(3):P or S(3):F

Safety(4) Patient Leakage Current

S(4):P or S(4):F

Safety(5) Patient Leakage Current with Mains

S(5):P or S(5):F