General Site Preparation Safety Information

2-2

Site Preparations

General Site Preparation Safety Information

Warning

Warning

To avoid contaminating or infecting personnel, the service environment or other

equipment, make sure that equipment which has been used before has
been appropriately disinfected and decontaminated.

Disconnect the monitor from the AC source by unplugging the power cable from the AC

power connector located on the rear of the monitor. The On/Standby
button does not disconnect the monitor from the AC mains supply

Accessory equipment connected to the monitor’s data interface must be certified

according to IEC Standard 60950 for data-processing equipment or
IEC Standard 60601-1 for electromedical equipment. All combinations
of equipment must be in compliance with IEC Standard 60601-1-1
systems requirements.

Explosion Hazard. Equipment not suitable for use in the presence of a flammable

anaesthetic mixture with air, or with oxygen or nitrous oxide.

Make sure that you have read all applicable instructions before attempting to install the

wall mount.

Wall mounts that are intended to support monitors must be capable of supporting 4

times the weight of the monitor when properly installed.

Do not mount any portion of a monitoring instrument over a patient’s bed.

Do not exceed the maximum rated load specified for each wall mount.

Ensure that no electrical wiring, utilities, or piping interfere with the selected wall

mounting location.

Note

Do not return sensors, patient cables, NBP tubing and cuff, or the power cord.