Objectives, Concepts, Functionality assurance – Philips 862478 User Manual

Page 53: Preventative maintenance, Performance and safety tests, Objectives -3 concepts -3

Advertising
background image

Objectives

Testing the Monitor

4-3

Objectives

In order to meet this chapter’s objectives, you should be able to test the monitor through the
following types of tests:

Performance Assurance Check List

Accuracy, Calibration and Performance Procedures

Patient Safety Checks

For inspection procedures; preventative maintenance procedures; cleaning procedures; and
battery maintenance, refer to this chapter and Chapter 2, “Site Preparations”.

Concepts

Functionality

Assurance

This refers to the combined Performance Assurance Test and Functionality Testing
Procedures found in this chapter. These tests verify correct monitor functionality in general
terms.

Preventative

Maintenance

Preventative Maintenance refers specifically to the service calibration tests required to ensure
the monitor measurement results are accurate. When authorized Philips personnel service the
monitor, they report these results back to Philips. The collected data forms a database to be
used in product development. These specific tests are required for the NBP parameter. It is not
necessary for hospital personnel to report results.

Performance

and Safety

Tests

This concept refers to all the remaining accuracy and performance tests available on the
monitor, including safety tests and checks for the monitor.