Diagnostic Mode

5-8

Configuring the Power-up Defaults Menu

System A/D

Values

The System A/D screen displays the current value of each analog-to-digital (A/D) channel in
volts. Some of the channels are for AC-coupled signals (such as ECG input), so the numbers
on the screen are constantly changing when an input signal is present. These AC-coupled
values are shown to give an indication as to whether basic functionality of the channel is
present, but no significance can be derived from the values of the numbers displayed.

The Primary and Secondary Status messages from the SpO

2

module are displayed and

updated at the rate of about once per second. Presence of the correct SpO

2

message indicates

that, at a basic level, communication between the SpO

2

module and the main monitor

processor is working correctly. None of the displayed values can be changed or reset in this
screen. When in the System A/D screen, the Return option is always highlighted. Press the
wheel to return to the Diagnostic Menu. Rotating the wheel while in the System A/D screen
has no effect. The A/D channel designators are shown in the following table.

DIAGNOSTIC MENU

SYSTEM A/D VALUES

1. 12.

2.

13.
3. 14.

4. 15.

5. 16.
6. 17.

7. 18.
8. 19.

9. 20.

10. 21.
11.

Return

SpO2 S1 SpO2 Alive SpO2 S2 No Sensor