Restoring Factory Settings

5-14

Configuring the Power-up Defaults Menu

Restoring Factory Settings

Caution

In addition to restoring factory defaults, this procedure clears the contents of trend memory.

The following technique can be used to restore the monitor’s power-up default settings which
were originally configured in the factory.

Note

Before performing the first step, read this procedure COMPLETELY.

Step

Action

1

Ensure that the monitor is powered Off.

2

Simultaneously press the Volume and Contrast buttons on the monitor’s front
panel.

3

Still pressing the Volume and Contrast buttons, press the On/Standby button to
turn the monitor On.

4

Continue to press the Volume and Contrast buttons until the power-up
diagnostic sequence is complete. When the main monitoring screen displays,
release the two buttons.

5

Select your language from the following screen, then select YES to place the
monitor into that language. After language is selected, the main monitoring
screen displays.