Error code categories, Error code categories -12 – Philips 862478 User Manual

Page 122

Advertising
background image

Part 3 Using Support Functions

6-12

Troubleshooting

Error Code

Categories

As a reference, the following table lists the general categories for other error codes. The error
code categories are shown only in hexadecimal format.

64

The SpO

2

module is sending an

error message to the host CPU.

1. Ensure PicoSat module is

properly connected.

2. Replace the PicoSat module.
3. If problem persists, replace main

PCB module.

?

The CO

2

module is sending an

error message to the host CPU.

1. Ensure Oridion module is

properly connected.

2. Rplace the Oridion module.
3. If problem persists, replace main

PCB module.

65, 66

PicoSat SpO

2

module has detected

an error during initialization.

1. Replace PicoSat module.
2. If problem persists, replace main

PCB module.

6E-71

PicoSat SpO

2

module has detected

an error on its serial port.

1. Replace PicoSat module.
2. If problem persists, replace main

PCB module.

Hex Code

Explanation

Recommended Action

Hexadecimal Code

Explanation

500xxx

internal user interface error

501xxx

remove serial port error

502xxx

date and time error

503xxx

NBP error

504xxx

front end error

505xxx

alarm error

506xxx

audio error

507xxx

recorder error

508xxx

trend error

509xxx

flash memory data error

50axxx

SpO

2

error

50bxxx

ECG error

50cxxx

power-down task error

50dxxx

on-board diagnostic error

50exxx

power monitor error

50fxxx

temperature measurement error

510xxx

internal user interface error

511xxx

error handling error

513xxx

serial driver error

Advertising