Co2 -9 – Philips 862478 User Manual

Page 181

Advertising
background image

Measuring and Displaying Parameters

Specifications

12-9

CO

2

Complies with EN 864/ISO 9918

Parameter

Specification

Range

0 to 98 mmHg (0 to 13 kPa), or 13%
CO

2

whichever is lower.

Flow Rate

50 ±7.5 ml/min

Warm-up Time

180 seconds max.; typically 30 seconds

Gas Sampling Delay Time

2.7 seconds

Resolution

Numeric

1.0 mmHg (0.1 kPa)

Wave

0.1 mmHg (0.01 kPa)

Rise Time

190 ms for adult mode (measured with
sample line for humidified ventilation
and airway adapter for adult)

Calibration Interval

4,000 operating hours

Auto Zero Interval

< 2% duty cycle. Increases if the
temperature change by 8

o

C from the

previous Auto-Zero interval, or the
pressure changes by 45 mmHg.

Leak Tightness

< 40 mbar/min with 30% vacuum

Accuracy

a

a. Accuracy applies for respiration rate up to 80 b/min. For respiration rate above 80 b/min, the

accuracy complies with EN 864/ISO 9918 (4 mmHg or ±12% of indicated value, whichever
is greater) for End-tidal CO

2

values exceeding 19 mmHg. The above accuracy is maintained

to within 4% for the following gas mixtures: CO

2

: 0 - 13%; N

2

: 0 - 97.5%; O

2

: 0 - 100%;

N

2

O: 0 - 80%; H

2

O: Dry to saturated.

0 - 20 minutes

0 to 38 mmHg ±4 mmHg; 38 to 99
mmHg ±12% of indicated value

> 20 minutes

0 to 38 mmHg ±2 mmHg
38 to 99 mmHg ±5% of indicated value
+ 0.08% for every 1.0 mmHg above 40
mmHg.