Documenting nbp test results, Documenting nbp test results -15 – Philips 862478 User Manual

Page 65

Advertising
background image

Preventative Maintenance Tests

Testing the Monitor

4-15

Documenting

NBP Test

Results

The following table lists the tests that should be documented and summarizes how to
document the NBP test results.

Test

Expected Test Results

What to record on
service record

Pressure Transducer Accuracy Test

X1 = difference between the pressure on
the monitor and on the simulator

Difference less than or equal to 5
mmHg (0.7 kPa)

PN:P/X1/X2/X3/X4/X5

or

PN:F/X1/X2/X3/X4/X5

P = passed, F = failed

Pneumatic Leakage Test

X2 = difference between the pressure at
the start (P1) and after one minute (P2)

Difference less than or equal to 6
mmHg (0.8 kPa)

Inflation Rate Test

X3 = number of seconds for the monitor
to inflate

Rate between 1 and 6 seconds

Over-Pressure Test

X4 = peak value

Peak value between 270 and 330 mmHg
(36 kPa and 44 kPa)

Deflation Rate Test

X5 = mmHg (kPa) per second to deflate

1.8 mmHg/s to 4.8 mmhg/s