Performance Procedures

4-30

Testing the Monitor

Documenting

CO

2

Test

Results

The following table lists the CO

2

tests and summarizes how to document the test results.

Test

Expected Test Results

What to record on service record

Barometric Pressure Check

X1 = difference between the reference
pressure and the measured ambient
pressure displayed on the monitor
(-12 mmHg < X1 < +12 mmHg)

PCO

2

: P/X1/X2/X3/X4/X5/X6/X7

or

PCO

2

: F/X1/X2/X3/X4/X5/X6/X7

P = passed, F = failed

X1: xx (two digits)

X2: x.x

X3: x.x

X4: xxx

X5: x.x

X6: x.x

X7: x.x

Leakage Check parts 1 and 2

X2 = value of part 1 leakage check on
flowmeter (X2 </= 4.0 ml/min)

X3 = value of part 2 leakage check on
flowmeter (X3</= 4.0 ml/min)

Pump Check

X4 > 120 mmHg below the ambient
pressure

Flow Rate Check

X5 = CO

2

flow rate (42.5 ml/min <X5

<65 ml/min).

CO

2

Gas Measurement Calibration

Check

X6 = difference between measured
CO

2

value and calculated value, based

on 5% CO

2

cal. gas (X7<2.6 mmHg,

or 0.35 kPa)

Calibration Verification

X7 = difference between measured
CO

2

value and calculated value, based

on 10% CO

2

cal. gas.

(X7±{0.07 x value calculated})