Leakage check, Leakage check -32 – Philips 862478 User Manual

Page 82

Advertising
background image

Performance Procedures

4-32

Testing the Monitor

Leakage

Check

The Leakage Check consists of two parts:

Part 1 - Checking the tubing between the pump outlet and the

CO

2

module outlet.

Part 2 - Checking the tubing between the pump inlet and the FilterLine inlet.

Note

Check the flowmeter’s user guide for details on how to make a correct flow reading.

Part 1

Part 2

Step

Action

1

Check the ambient and cell pressure shown in the status line on the screen. The
cell pressure should be approximately 20 mmHg (2.7 kPa) lower than the
ambient pressure.

2

Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting
tube.

3

Block the CO

2

module outlet using your fingertip and observe the flowmeter

display. The value on the flowmeter (X2) should decrease to between 0 and 4 ml/
minute. If the value is not within the tolerance limits, there is a leakage between
the pump outlet and the CO

2

module gas outlet.

4

If a leakage is found in Step 3, open the CO

2

module and check the tubing

connections at the pump outlet and the module CO

2

gas outlet. If the connections

are good, then there is leakage in the module and the CO

2

module must be

exchanged.

Step

Action

1

Disconnect the flowmeter from the Part 1 setup and connect the flowmeter inlet
to the CO

2

module gas outlet.

2

Leave the FilterLine connected to the CO

2

module inlet.

3

Block the inlet of the FilterLine using your fingertip and observe the flowmeter
display. The value on the flowmeter (X3) should decrease to between 0 and 4 ml/
minute. If the value is within the tolerance limits, there are no leakages and the
leakage check is completed. You can proceed to the “Pump Check”.

4

If the value is not within the tolerance limits, there is a leakage between the
FilterLine inlet and the pump inlet.

5

Check the FilterLine connections and open the monitor to check the tubing
connections at the pump inlet and the CO

2

module gas inlet. If the connections

are good, try replacing the FilterLine and repeating the leakage check. If the
situation remains, there is a leakage in the tubing and the CO

2

module must be

exchanged.