Performance Procedures

4-38

Testing the Monitor

ECG Sync

Test

This test checks the performance of ECG synchronization between the monitor and a
defibrillator. The ECG sync performance test is required once every year and when the
monitor is repaired or when the monitor’s parts are replaced.

Tools Needed:

•

Dynatech Nevada MedSim 300 or equivalent simulator

•

Defib Sync Cable (M4820A) or Switchcraft 850

•

Switchcraft 750

Two sections are needed to complete the connection from the monitor to the simulator.

1. If a Defib Sync Cable is available, connect it the ECG Sync output on the monitor, then

to the Switchcraft 750.
or
If a Switchcraft 850 is available, connect it the ECG Sync output on the monitor, then
to the Switchcraft 750.

2. Connect the Switchcraft 750 to the SYNC/A PACE input on the simulator

Connect to the monitor

Connect to the simulator

Defib Sync Cable

Switchcraft 750

Switchcraft 850

Switchcraft 750

Step

Action

1

Connect the ECG leads and trunk cables between the monitor and the simulator
(as described above).

2

Turn on the monitor and the simulator

3

On the simulator main menu select DEFB, then CARD to set the simulator to
Defibrillator Tests/Cardioversion.

4

Press Start to begin the test. If the monitor is working properly, the following
sequence of text messages will appear on the simulator over a period of
approximately 3 seconds: "cardioversion: afib", "cardioversion: sync ok", and
"cardioversion: converted" (this message will persist on the simulator)

5

If the monitor isn’t working properly, messages such as the following may
appear: "cardioversion: vfib", "cardioversion: late".