Patient Safety Tests

Testing the Monitor

4-39

Patient Safety Tests

Philips safety tests meet the standards of, and are performed in accordance with IEC 60601-1,
Clause 19 (EN60601-1, Second Edition, 1988; Amendment 1, 1991-11, Amendment 2, 1995-
03).

The C3 patient monitor is a Class I device. It requires a protective earth (ground) wire. Keep
this in mind when performing the following test procedures.

There are two categories of safety tests:

Ground Integrity

Electrical Leakage

Ground

Integrity

Table 4-3. Ground Integrity

Test or “Inspection” to Perform

Expected Test Results

Protective Earth

See “Safety Test Diagram - Protective
Earth” on page 4-40.

With mains cable:

Maximum impedance = x <= 100 mOhms