Patient Safety Tests

Testing the Monitor

4-41

Electrical

Leakage

The following tests verify the electrical leakage of the monitor:

Earth Leakage Current

Enclosure Leakage Current

Patient Leakage Current

Patient Source Current, with Mains Voltage on the Applied Part

Warning

Warning

Perform all leakage tests any time the unit is opened.

Earth Leakage

Current

This test is in compliance with IEC 60601-1 (Earth Leakage Current). In locations where
mains voltage is 100-120 volts, the applied voltage is 132 volts. In locations where mains
voltage is 220-240 volts, the applied voltage is 264 volts. The applied AC frequency should
be the same as the local mains (50 or 60 Hz).

All measurements shall be made with the power switch in both On and Off positions.

1. Connect the monitor AC plug to the electrical safety analyzer as recommended by the

analyzer operating instructions.

2. Perform test as recommended by analyzer operating instructions.

Earth leakage current is measured under various conditions of the AC mains and protective
earth conductor. For each condition, the measured leakage current must not exceed that
indicated below.

Table 4-4. Earth Leakage Current Values

What to record on the service record:

S(2): P or S(2): F

Test Condition

Polarity

Allowable Leakage Current

Normal

Normal

300 µA

Reversed

300 µA

S.F.C

a

Open Supply

a. S.F.C = Single Fault Condition

Normal

1000 µA

Reversed

1000 µA