System specifications, Section 15, See section 15 system specifications for relay – YSI 5200 User Manual
Page 125

YSI Incorporated
123
5200 Recirculating System Monitor
Section 15. System Specifications
Temperature
Sensor type
10,000 ohm @ 25°C YSI Precision
™ thermistor
Range
0 to 45°C
32 to 113
o
F
Accuracy
± 0.2 °C
Resolution 0.1°C
Dissolved Oxygen
Sensor type
Membrane covered steady state polarographic
Range
0 to 500% air saturation
0 to 60 mg/L
Accuracy
±2% or ±0.2 mg/L, whichever is greater (0–200%)
±6% of reading (200–500%)
Resolution
0.1% or 0.01 mg/L (0–10 mg/L)
1% or 0.1 mg/L (10–60 mg/L)
Response time
98% of reading in 5 minutes
pH
Range
0.0 to 14.0 pH
Accuracy 0.1
pH
Resolution 0.01
pH
ORP
Range
-1000 mV to +1000 mV
Accuracy ±10
mV
Resolution 1
mV
Conductivity
Sensor type
4 electrode, Cell constant of 5.0 ±5%
Ranges
0 to 200 µS
0 to 2000 µS
0 to 20 mS
0 to 200 mS
Accuracy
0 µS to 100 mS, ±0.5% of range
100 mS to 200 mS, ±1.0% of reading
Resolution
1.0 µS, 0.1mS / Range dependent