Alarm logic flow charts, Section 7, 1 alarm logic flow charts – YSI 5200 User Manual

Page 75: Ysi incorporated, General alarming

Advertising
background image

YSI Incorporated

73

5200 Recirculating System Monitor

Section 7. Alarm/Pager

7.1 Alarm Logic Flow Charts

Is Parameter

outside alarm

values

Is alarm for this

parameter enabled

Is Alarm holdoff

enabled

Is Gerneral Alarm

enabled

Count down

holdoff time

Is Pager

enabled

Send
Page

Yes

No

Yes

No

Yes

No

No

Yes

Stop

Stop

No

Yes

Activate
General

Alarm

relay

Parameter

data

Activate

beeper

and Red

LED for

parameter

Email

General Alarming

Is individual

parameter alarm

relay assigned

Yes

Activate

individual

parameter

Alarm

relay

No

Has Alarm

been cleared

Wait for pager

retransmit time

Stop

No

Yes

Advertising