Dissolved oxygen calibration, Do calibration in % saturation, L. see section 6.2.2 dissolved oxygen calibration – YSI 5200 User Manual

Page 65: 2 dissolved oxygen calibration

Advertising
background image

YSI Incorporated

63

5200 Recirculating System Monitor

6.2.2 Dissolved

Oxygen

Calibration

This procedure calibrates dissolved oxygen. Calibrating any one option (% or mg/L) automatically
calibrates the other.

Salininty 0-45ppt

mg/L

%Sat

Display Actual Value % (CAL Value via Math)

Scrolling Text*1

Display Actual Value mg/L

Scrolling Text*2

Calibration

Complete

Yes

Calibration

If Cond turned

Off Display

Question

No

Alt 100 X 100 Ft

Calibration

Complete

Get UserValue

Figure 6-1 DO Calibration Menu

DO Calibration in % Saturation

1. Go to the DO calibrate screen as described in Section 6.2.1 Accessing the Calibrate Screen.

2. If your probe module contains a conductivity probe and it is Enabled, skip to Step 3. If your probe

module does not contain a conductivity probe (or the conductivity probe is not Enabled), enter
the salinity of the sample stream.

NOTE: The salinity compensation range for DO is 0–45 ppt.