5200 Recirculating System Monitor

Section 16. Factory Default Settings

Setting

Factory Default Value

Your Value

Parameters—Auxiliary

Aux Input 1

Disabled

Aux Input 1 Control

Close

Aux Input 1 Hold Off

15 Sec

Aux Input 1 Alarm

No

Aux Input 1 Relay

Disabled

Aux Input 2

Disabled

Aux Input 2 Mode

Digital

Aux Input 2 Control

Close

Aux Input 2 Hold Off

15 Sec

Aux Input 2 Alarm

No

Aux Input 2 Alarm Low

-2000

Aux Input 2 Alarm High

+2000

Aux Input 2 Relay

Disabled

Aux Input 2 Range

0–1V

Aux Input 2 Format

0

Aux Input 2 Min

-5000

Aux Input 2 Max

+5000

Aux Input 2 Suffix

(blank)

Aux Input 2 Label

(blank)

Parameters—Conductivity

Conductivity

Enabled

Conductivity Range

Auto

Conductivity Temp Comp

1.91

Conductivity Set Point

1,600 uS

Conductivity High Control

1,800 uS

Conductivity High Control Relay

Disabled

Conductivity Low Control

1,400 uS

Conductivity Low Control Relay

Disabled

Conductivity High Alarm

1,900 uS

Conductivity High Alarm Relay

Disabled

Conductivity Low Alarm

1,300 us

Conductivity Low Alarm Relay

Disabled

Conductivity Dose Up Time

30 Seconds

Conductivity Dose Up Wait

15 Minutes

Conductivity Dose Down Time

30 Seconds

Conductivity Dose Down Wait

15 Minutes