CAUTION: Do not touch the membrane surface.

8. Screw the membrane cap onto the sensor moderately tight. A small amount of electrolyte should

overflow.

9. IMPORTANT: To prevent the formation of a bulge in the membrane (which can slow sensor

response), partially unscrew the membrane cap then retighten.

10. Screw the probe sensor guard on moderately tight.

4.7 Calibration/Storage Cup (5562 only)

The YSI 5562 Probe module is supplied with a convenient calibration/storage cup. This cup is an ideal
container for calibration of the different sensors, minimizing the amount of solution needed. Refer to
Section 6 Calibration.

4.7.1

Calibration/Storage Cup Installation

1. Remove probe sensor guard, if already installed.

2. Ensure that an o-ring is installed in the o-ring groove on the threaded end of the probe module

body.

3. Screw the calibration/storage cup on the threaded end of the probe module and securely tighten.

NOTE: Do not over tighten as this could cause damage to the threaded portions.

Figure 4-9 Calibration/Storage Cup Installation

4.8 Instrument/Cable

Connection

Attach the probe cable to the instrument as follows:

1. Line up the pins and guides on the cable with the holes and indentations on the cable connector at

the bottom of the YSI 5200 instrument.