Table 1: Solubility of Oxygen (mg/L) in Water Exposed to Water-Saturated Air at 760 mm Hg Pressure.

Salinity = Measure of quantity of dissolved salts in water.

Chlorinity = Measure of chloride content, by mass, of water.

S(

0

/

00

) = 1.80655 x Chlorinity (

0

/

00

)

0

/

00

= ppt

Temp

o

C

Chlorinity 0

Salinity: 0

5.0

ppt

9.0

ppt

10.0

ppt

18.1

ppt

15.0

ppt

27.1

ppt

20.0

ppt

36.1

ppt

25.0 ppt

45.2 ppt

0.0

14.62

13.73

12.89

12.10

11.36

10.66

1.0 14.22

13.36

12.55 11.78 11.07 10.39

2.0 13.83

13.00 12.22 11.48 10.79 10.14

3.0 13.46

12.66 11.91 11.20 10.53 9.90

4.0 13.11

12.34 11.61 10.92 10.27 9.66

5.0 12.77

12.02 11.32 10.66 10.03 9.44

6.0 12.45

11.73 11.05 10.40 9.80 9.23

7.0 12.14

11.44 10.78 10.16 9.58 9.02

8.0 11.84

11.17 10.53 9.93 9.36 8.83

9.0 11.56

10.91 10.29 9.71 9.16 8.64

10.0 11.29

10.66 10.06 9.49 8.96 8.45

11.0 11.03

10.42 9.84 9.29 8.77 8.28

12.0 10.78

10.18 9.62 9.09 8.59 8.11

13.0 10.54

9.96 9.42 8.90 8.41 7.95

14.0 10.31

9.75 9.22 8.72 8.24 7.79

15.0 10.08

9.54 9.03 8.54 8.08 7.64

16.0 9.87

9.34 8.84 8.37 7.92 7.50

17.0 9.67

9.15 8.67 8.21 7.77 7.36

18.0 9.47

8.97 8.50 8.05 7.62 7.22

19.0 9.28

8.79 8.33 7.90 7.48 7.09

20.0 9.09

8.62 8.17 7.75 7.35 6.96