Parameter, Do/temp/ph low, Do/temp/ph high – YSI 5200 User Manual

Page 100: Do/temp/ph change, Setup menu, Relay, Units, Test, Calibrate feeder, Setup menu relay

Advertising
background image

YSI Incorporated

98

5200 Recirculating System Monitor

Day1 Day

2 Day

3 Day

4 Day

5 Day

6 Day

7

Biomass

1000 1020 1040 1061 1082 1104 1126

Feed

%

(of)

Biomass 3% 3% 3% 3% 3% 3% 3%

Daily

Feed

Amount

30 30.6 31.2 31.8 32.5 33.1 33.8

FCR

Factor

1.5 1.5 1.5 1.5 1.5 1.5 1.5

1/FCR

x

Feed

Amount

20 20.4 20.8 21.2 21.6 22.1 22.5

New

Biomass

1020 1040 1061 1082 1104 1126 1149

Parameter

Changes in water parameter conditions will affect the amount of food fish eat and thus will affect fish
growth. With the parameter feature enabled, the feed timer system will verify that pH, temperature and
DO values are within user selected ranges. This verification is done immediately prior to starting each
feed cycle duration. If the values are outside these ranges, food weight amount is reduced by a user
selected percentage.

DO/Temp/pH Low

Enter the lower limit of the parameter range where no feed compensation will take place.

DO/Temp/pH High

Enter the high limit of the parameter range where no feed compensation will take place.

DO/Temp/pH Change

Enter the percent reduction in feed weight that will occur for each 1/10 unit of measure the parameter is
out of range. Results from all out of range parameters are added.

Setup Menu

Relay

Assign a relay address to the feeder.

NOTE: Feed calibration cannot be performed if the Relay is disabled.

Units

Select lb (pound), g (gram), kg (kilogram), or oz (ounce) as the unit of weight the Feed Timer System
uses for the calibration/test sample amount and to the total daily feed amount.

NOTE: If the units are changed, the feeder must be recalibrated.

Test

Calibrating (Testing) the feeder is essential for proper operation of the Feed Timer System. Calibration
will determine the rate at which food is dispensed from a feeder. This information is used to determine
feed cycle duration(s).

NOTE: A relay must be assigned in order to enter the Calibrate Feeder menu.

Calibrate Feeder

Enter the amount of time the feeder relay will be energized for testing; 001–500 seconds. The relay will
energize after the last digit is entered. After the timer counts down to zero, the feed relay deenergizes and
the Measured Weight screen is displayed.